Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection

Permanent URI for this collectionhttps://hdl.handle.net/11147/7148

Browse

Search Results

Now showing 1 - 2 of 2
  • Article
    Citation - WoS: 2
    Citation - Scopus: 3
    Design and Manufacturing of a Hip Joint Motion Simulator With a Novel Modular Design Approach
    (Springer, 2023) Torabnia, Shams; Mihçin, Şenay; Lazoğlu, İsmail
    The study is aimed to develop a hip joint wear simulator using a modular design approach to help experimentally monitor and control critical wear parameters to validate in-silico wear models. The proper control and application of wear parameters such as the range of motion, and the applied force values while estimating the lost material due to wear are essential for thorough analysis of wear phenomena for artificial joints. The simulator's dynamics were first modeled, then dynamic loading data was used to calculate the forces, which were further used for topology optimization to reduce the forces acting on each joint. The reduction of the link weights, connected to the actuators, intends to improve the quality of motion transferred to the femoral head. The modular design approach enables topology-optimized geometry, associated gravitational and dynamic forces, resulting in a cost-effective, energy-efficient product. Moreover, this design allows integration of the subject specific data by allowing different boundary conditions following the requirements of industry 5.0. Overall, the in-vitro motion stimulations of the hip-joint prosthesis and the modular design approach used in the study might help improve the accuracy and the effectiveness of wear simulations, which could lead into the development of better and longer-lasting joint prostheses for all. The subject-specific and society-based daily life data implemented as boundary conditions enable inclusion of the personalized effects. Next, with the results of the simulator, CEN Workshop Agreement (CWA) application is intended to cover the personalized effects for previously excluded populations, providing solution to inclusive design for all.
  • Conference Object
    Citation - WoS: 1
    Citation - Scopus: 2
    Parametric Analysis for the Design of Hip Joint Replacement Simulators
    (IEEE, 2021) Torabnia, Shams; Mihçin, Şenay; Lazoğlu, İsmail
    The simulation of wear, between the components of artificial hip joint implants, is a complicated problem that does not have a robust analytical answer yet. Many studies have been conducted to predict the wear between the femur head and the acetabular cup, as the debris generated due to the wear might produce adverse effects after the surgery. Hip joint simulators provide a means to quantify the amount of wear in preclinical settings, as an in vitro method. However, this brings some other challenges in terms of bio-fidelity. The simulators use force and range of motion data as input and provide wear information as an output. For this reason, it is important to be able to simulate the realistic conditions, by the proper transmission of force and position controlling of the components. Many studies performed on wear simulators but none of them worked on the machine parameters such as power consumption and sensitivity to external inputs in detail. In this study, we perform a sensitivity analysis of the factors affecting the forces acting on the femur head. In silico simulations were performed by changing the values of acting force, friction coefficient, and radius of femur head to understand the effects of each parameter on the frictional moment of the joint. These analyses demonstrate the importance of using correct parameters while designing simulators, which accept flexible boundary conditions. The architecture of the hip simulator was also investigated for the first time. The results are expected to pave the way for improving the bio-fidelity of the simulators in the field of biomechanics. © 2021 IEEE.