Gravity Compensation of a 2r1t Mechanism With Remote Center of Motion for Minimally Invasive Transnasal Surgery Applications [article]

Loading...

Date

Authors

Ayit, Orhan
Kiper, Gökhan
Dede, Mehmet İsmet Can

Journal Title

Journal ISSN

Volume Title

Open Access Color

Green Open Access

No

OpenAIRE Downloads

OpenAIRE Views

Publicly Funded

No
Impulse
Top 10%
Influence
Average
Popularity
Top 10%

relationships.isProjectOf

relationships.isJournalIssueOf

Abstract

This work addresses the gravity balancing of a 2R1T (2 rotations – 1 translation) mechanism with remote center of motion. A previously developed balancing solution is modified and applied to a prototype and test results are presented. The mechanism is an endoscope holder for minimally invasive transnasal pituitary gland surgery. In this surgery, the endoscope is inserted through a nostril of the patient through a natural path to the pituitary gland. During the surgery, it is vital for the manipulator to be statically balanced so that in case of a motor failure, the patient is protected against any harmful motion of the endoscope. Additionally, static balancing takes the gravitational load from the actuators and hence facilitates the control of the mechanism. The mechanism is a 2URRR-URR type parallel manipulator with three legs. The payload mass is distributed to the legs on the sides. By using counter-masses for two links of each leg, the center of mass of each leg is lumped on the proximal link which simplifies the problem to balancing of a two degree-of-freedom inverted pendulum. The two proximal links with the lumped mass are statically balanced via springs. Dynamic simulations indicate that when the mechanism is statically balanced, generated actuator torques are reduced by 93.5%. Finally, the balancing solution is implemented on the prototype of the manipulator. The tests indicate that the manipulator is statically balanced within its task space when the actuators are disconnected. When the actuators are connected, the torque requirements decrease by about 37.8% with balancing.

Description

Keywords

Surgery applications, Robotics in medicine, Computer-assisted surgery, Parallel manipulators

Fields of Science

0209 industrial biotechnology, 02 engineering and technology

Citation

WoS Q

Scopus Q

OpenCitations Logo
OpenCitations Citation Count
7

Source

Volume

41

Issue

3

Start Page

807

End Page

820
PlumX Metrics
Citations

Scopus : 8

Captures

Mendeley Readers : 4

SCOPUS™ Citations

7

checked on Apr 27, 2026

Web of Science™ Citations

7

checked on Apr 27, 2026

Page Views

6635

checked on Apr 27, 2026

Downloads

3428

checked on Apr 27, 2026

Google Scholar Logo
Google Scholar™
OpenAlex Logo
OpenAlex FWCI
0.77698159

Sustainable Development Goals