Almost all external radiotherapy planning systems use dose profile data. There are various methods to measure these profiles. Direct measurements with semiconductor detectors are widely used. But this method takes lots of time especially at linear accelerators. We developed a simple method for this purpose. Dose profiles are measured with using radiographic films and verification films. Suitable modifications are done with derived sensitometric curve. Results are compared with original profiles. For the Co60, this method is found as an alternative to other methods with acceptable accuracy (%±2)
Description
Keywords
Profile, OAR, Radiotherapy, Film, OAR, Radiotherapy, Profile, Film
Fields of Science
03 medical and health sciences, 0302 clinical medicine, 0202 electrical engineering, electronic engineering, information engineering, 02 engineering and technology
Citation
Gökçe, T. C., Aytaç, S., and Gökçe, T. (2001, October). Determination of dose profile data with film dosimetry. Paper presented at the Proceedings of the 23rd Annual EMBS International Conference, İstanbul, Turkey, (pp. 3923-3924). Piscataway, N.J. : Institute of Electrical and Electronics Engineers.